PRODUCT CLASSIFICATION
產(chǎn)品分類工業(yè)物聯(lián)網(wǎng)監(jiān)控系統(tǒng)
我司成功開發(fā)中英文版本土壤無核密度儀
HO-4114土壤無核密度儀/無核密濕度儀 簡稱:無核儀
This is a general user manual for Electrical Density Gauge. It contains a description of the EDG and its components, and guidelines for use and calibration. It is strongly recommended that you review this manual carefully before using EDG. The information in this manual has been carefully checked and is believed to be entirely reliable. However, no responsibility is assumed for inaccuracies. Furthermore, EDG, LLC reserves the right to make changes to products herein described without prior notice, so as to improve reliability,function, or design.
TABLE OF CONTENTS
Introduction to the EDG Process ………… .…….. ……. ………….. 3
Components 5
The process steps of HO-4114 8
Introduction of HO-4114 Functions in MainFrm 11
Simple introduction of submenu of MainFrm: 11
(1).Create Soil Modal file SED-XX and STD-XX. 12
Introduction
Your Electrical Density Gauge (EDG) is capable of providing accurate readings of soil density, moisture content, and percent compaction on soils typically used for roads and foundations. The precision by which the EDG can determine these physical parameters depends highly upon you, the operator. Reading and understanding the contents of this manual is a very important step toward utilizing EDG to its maximum potential.
The following will give you an overview of the EDG process and help you understand the more detailed steps found later in the manual.
Is EDG Safe To Use?
Absoluy! There is far less measuring energy present when using EDG than when you are using a cell phone.
Introduction to the EDG Process
The EDG process introduces some new terminology with which you need to be familiar.
A Soil Model is a combination of electrical soil data and user-entered physical soil data. The Soil Model provides the basis for testing a specific soil type within an area of similar soil characteristics. To make a Soil Model, several electrical signatures are acquired using the EDG at different Test Spots (the Test Spots for a Soil Model are called Soil Tests.) For each Test Spot that an electrical signature is acquired, a Sand Cone test is also performed to obtain physical wet density, dry density, and percent moisture information. A proctor test can optionally be performed to determine the maximum dry density. The physical information determined by the Sand Cone and Proctor Mold laboratory tests are entered into the EDG unit ,thereby tying electrical signature information to physical data.
The Soil Model can be thought of as a way of “calibrating” the EDG to the particular soil type. At least three Test Spots (Soil Tests) are required to create a Soil Model; more Test Spots make for a more representative Soil Model. Further, Test Spots that exhibit a broad range of density and moisture content make for more accurate Soil Models.
Once the Soil Model has been created, the EDG is ready for testing at aJob Site. A Job Site is an area of similar soil characteristics in which the user wishes to determine the physical characteristics of the soil. A single Soil Model is generally created for each different Job Site.
Once a Soil Model has been created that accuray models the Job Site,the soil characteristics of any Test Spot within the Job Site can be accuray and quickly determined by the EDG by performing a Field Test. A Field Test is electrical soil data at a particular Test Spot that is fed into the Soil Model to produce physical soil characteristics.
1. EDG Console/Computer (1)
2. Electronic Soil Measurement Sensor (1)
3. Calibration Check Unit (1)
4. Dart Template (1)
5. Soil Darts (4)
6. Temperature Probe (1)
7. Hammer (1)
8. Battery Charger (1)
9. Cables/Clips for Soil Measurement Sensor (1)
10. Break-Out Cable (1)
11.HO-4114Manual(1)
12. Software (1)
Components:
Soil Darts: To get electrical data from the soil, 4 Soil Darts are pounded into the ground with a hammer. A template is provided to accuray position the Soil Darts. The Soil Darts are machined from stainless steel,and are expected to last many years. If the conical end becomes flattened from hitting too many rocks, either file or grind off any mushrooming at the point to maintain a conical shape to the Dart. If the top end of the Dart becomes damaged or starts to crack from excessive hammering, file off the damaged part to avoid having pieces of steel flying off when driving in the Dart. Always wear safety glasses when hammering in the Soil Darts.
Electrical Soil Measurement Sensor: The Soil Sensor gathers electrical information about the soil under test and transfers the information to the EDG computer. The Soil Sensor is connected to the Soil Darts using cables and clips. The Soil Sensor connects to the EDG unit using the left panel receptacle.
The Soil Sensor requires no maintenance or adjustment, so there is no need to open the enclosure. Keep it clean and dry when using it. It is an electronic circuit, so treat it with the same respect you use with your cell phone. If the Sensor Pins become loose, hold the pin with pliers while tightening the nut
against the enclosure.
Temperature Probe :The temperature probe attaches to the left panel receptacle. When performing EDG testing, the temperature probe should be placed into the soil. The soil temperature is an important variable in accurate EDG testing. The EDG computer uses the temperature information to perform temperature compensation, resulting in more accurate EDG results. The temperature probe must be connected when creating Soil
Models; use of the temperature probe is optional when performing Field Testing. If the temperature probe is not used during Field Testing, the EDG results cannot be temperature compensated.
The Break-Out Cable: The Break-Out Cable provides a 9-pin Data Cable connector for connection to a host PC for use in Data Sharing mode.
EDG Battery/Battery Charger: The EDG battery is rated at a capacity of 4.0 Amp-Hours. This is enough energy for continuous operation for over 8 hours The Battery Charger that is provided with EDG will charge the battery in approximay 12 hours.Therefore, EDG should be placed on charge each day after use, and be allowed to charge overnight.
EDG Console/Computer: The EDG unit contains a computer that applies mathematical formulas to determine the physical characteristics of the soil. The computer also contains memory that remains intact after the EDG has been powered off; this memory is used to save Soil Models and Job Site data. See Appendix A for information on EDG memory capacity. The EDG console has an LCD display that is used to provide visual information to the user and a keypad that allows the user to key information into the EDG unit.
概述:土壤無核密度儀/無核密濕度儀,主機采用原裝美國進口機芯、國內(nèi)組裝,保證測量精度、中英文菜單顯示,操作簡單易學。
土壤無核密度儀/無核密濕度儀(EDG)是一種可替代核子密度儀,測量路基和地基壓實土壤之物理特性的無核測量儀。它是一種以電池驅(qū)動的便攜式儀器,可以在世界各地任何地方使用,而不會帶來核安全等相關問題。EDG可測量并顯示干/濕密度、含水量,以及壓實百分數(shù)。EDG使用簡便,可以作為建筑用輔助設備來監(jiān)控每日的壓實操作,其的工作性能和準確的測量結(jié)果可以與傳統(tǒng)測量法,如核子密度測量法,灌砂法,以及干燥法相媲美。EDG利用點對點的無線電頻率來確保度。在插入土壤的電極之間進行直接測量,并通過*覆蓋電極的土壤來確保正確測量。使用EDG測量時不必擔心土壤是否均勻,也無須依靠無線電頻率的“電波”從表面滲透入土壤材料。對于操作者而言,EDG的優(yōu)勢還在于它的操作簡單易學,不需要持有特殊資格證明或經(jīng)過專業(yè)培訓的技師,任何人都可以成為培訓對象。此外EDG測量快速,測量間隔距離更大,估算更*。
HO-4114土壤無核密度儀/無核密濕度儀測量原理:
無核密濕度儀(EDG)通過電極之間的無線電高頻率來測量壓實土壤材料的介電性和密度,并將測量得到的介電性與“土壤模塊”做比較。“土壤模塊”是土壤類型的特定標準,它有一組預先測量得到的特定介電性,這組介電性代表了一系列的密度值和濕度值。野外測量結(jié)束后,利用數(shù)學運算法則會計算出干/濕密度、含水量,以及壓實百分數(shù)。隨附的溫度探測器還可以增加測量的性。
HO-4114土壤無核密度儀/無核密濕度儀用途:
用于以下范疇壓實材料的密度、濕度測量:
筑堤和填充物;
地下道,地下室和地基;
回填和溝渠;垃圾;
HO-4114土壤無核密度儀/無核密濕度儀特點:
1.可代替核子密度測量法,灌砂法,以及干燥法。無核源!
2.不再需要核子儀器所必需的各種許可證。
3.操作員不需要專門的培訓/證書和輻射監(jiān)測。
4.存儲與運輸要求沒有特殊要求。
5.用戶界面良好、簡單易學、使用方便、性價比高,適合小組中每個人員使用。
6.快速,可靠,輕便、牢固、,重復性好的實時讀數(shù)。
7.采用高質(zhì)可靠的點對點RF技術(shù)
8.2-3分鐘顯示測量結(jié)果
9.確保建筑物,橋梁,道路和大壩等工程結(jié)構(gòu)的基礎牢固。
10.內(nèi)置中英文菜單,方便國內(nèi)外施工人員操作。
HO-4114土壤無核密度儀/無核密濕度儀技術(shù)指標:
濕度范圍:標準的壓實土壤現(xiàn)場范圍
干密度精度:標準測試的3%以內(nèi)
濕量范圍:標準的壓實土壤現(xiàn)場范圍
濕量精度:標準測試的2%以內(nèi)
操作環(huán)境溫度:0-50oC
操作環(huán)境濕度:5-90%
電源:12V,4.0Ah鉛酸鉻可充電電池
電池壽命:*充電可用約8小時
尺寸:控制臺:343mmx305mmx152mm
凈重:控制臺:11lbs.(5kg);附件:4lbs.(1.8kg)
HO-4114土壤無核密度儀/無核密濕度儀使用方法:
在野外現(xiàn)場,依據(jù)模板用錘子將錐形電極或標塊以十字式樣插入土中。在兩個錐形標塊的底角之間進行4個點對點的測量并將電特性平均處理。用戶可選擇用公制SI單位或英制單位顯示密度和濕度值。標槍插入的深度直接影響測量深度。此外,標塊的錐度還通過每一次錘擊與土壤保持正接觸。標塊的插入深度決定了測量深度—有不同長度可選。EDG顯示屏可顯示菜單操作步驟,保存測量結(jié)果。
關鍵詞:土壤無核密濕度儀/無核密濕度儀/土壤密濕度儀/無核密度儀/無核土壤密度濕度儀/土壤無核密度儀/無核土壤密度儀
郵箱:bjhoyq@163.com
傳真:010-67051434
地址:北京市朝陽區(qū)高碑店鄉(xiāng)北花園村6號(近韓國慕色攝影)